70%

Wallach Cryosurgical Tip – T-0219 Microderm Fashion

$145.80

Wallach Cryosurgical Tip – T-0219 Microderm

The Wallach Cryosurgical Tip – T-0219 Microderm (#900200AA) is utilized by the popular LL100, LLCO2, and WA1000B Cryosurgical Systems, and offers the ultimate in safety and convenience. The T-0219 Microderm tip measures 2 mm in diameter and 19 mm in protrusion.

Warranty: 90 Days.

A disposable shield can be removed from cryosurgical tip and is discarded after each use, which allows the cryosurgical tip to be sterilized. Avoid soaking the tips, which may lead to problems caused by oversoaking such as tip corrosion. Shields are renewed with each procedure, which insulate tips and assure proper attachment of tip by serving as a positive stop. 

Wallach Cryosurgical Tip Setup:

  • Remove disposable shield from cryotip and discard.
  • Plug cryotip before cleaning and sterilizing.
  • Scrub tip to remove blood and particles.
  • Rinse with water.
  • Steam sterilize cryotip.
  • Follow manufacturer’s directions for appropriate sterilization.
  • Attach new disposable plastic shield to the cryotip.
  • Cryotip is now ready to use.

Literature

Wallach Cryosurgical Tip – T-0219 Microderm Literature

Wallach LL100 Two-Trigger Cryosurgical Unit Directions for Use

Wallach LLCO2 Cryosurgical System Directions for Use

Wallach Cryosurgical Systems Sell Sheet

Wallach Reusable Sterilizable Cryotips and Disposable Plastic Shields Instructions for Use

Wallach Cryosurgical Tip - T-0219 Microderm